Medical Electrode Sheet

Vmanx Custom Medical Electrocardiographic Electrode Sheet

Reflect the excitation or inhibition state of the cerebral cortex, accurately provide detection and evaluation of ECG consciousness state, and meet clinical needs

Product Specification:
  • AC Impedance range : 0-20KΩ
  • Accuracy : ≤10KΩ 1%±10Ω, >10KΩ 3%
  • Test time : >60s
  • DC offset voltage Range : ±1200mV
  • Accuracy: ≤100mV1%±0.5 mV, >100mV1% :
  • Test time : 60-90s
INQUIRY NOW
Product Details

 

Electrical parameters of ECG electrode slices
Test items Excitation conditions Range Accuracy Test time
Impedance Sinusoidal alternating current
100uAp p - 5%
10Hz+10%
0-20KΩ ≤1KΩ 2%±10Ω
 >1KΩ  3%
>60s
Imbalance DC input resistance
≥10MΩ
±1200mV ≤100mV 1%±10Ω 
>100mV  2%
60-90s
Noise RC bandpass
0.15-100Hz
0-2000µV(P-P) ≤150µV ±7.5µV
>150µV  5%
60-360s
Current DC current 200nA±5% ±1200mV ≤100mV 1%±1mV 
>100mV  2%
more than 8h
Defibrillation Capacitance 10µF±5%
Power 200V±2%
±1200mV ≤100mV 1%±1mV 
>100mV  2%
5、15、25、35s

AC Impedance:

Excitation conditions: sinusoidal alternating current 100μAp-p±5%, 10Hz±10%

Range: 0-20KΩ, accuracy: ≤10KΩ 1%±10Ω, >10KΩ 3%, test time: >60s

DC offset voltage:

Excitation conditions: DC input resistance ≥ 10MΩ

Range: ±1200mV, accuracy: ≤100mV1%±0.5 mV, >100mV1%, test time: 60-90s

Compound Offset Instability and Internal Noise:

Excitation conditions: RC bandpass 0.15-100Hz

Range: 0-2000μV(p-p), accuracy: ≤150μV ±7.5μV, >150μV5%, test time: 60-360s

Bias Current Tolerance:

Excitation conditions: DC current 200nA±5%

Range: ±1200mV, accuracy: ≤100mV1%±0.5mV, >100mV1%, test time: more than 8h

Defibrillation overload recovery:

Excitation conditions: Capacitance 10μF±5%, power 200V±2%

Range: ±1200mV, accuracy: ≤100mV1%±0.5mV, >100mV1%, test time: 5, 15, 25, 35s.

 

Characteristics of ECG electrodes

1. Reflect the excitation or inhibition state of the cerebral cortex, accurately provide detection and evaluation of ECG consciousness state, and meet clinical needs;

2. The brain electrode does not contain latex and uses imported conductive adhesive and high-quality 3M double-sided adhesive, with low impedance and good adhesion;

3. Through biocompatibility testing, there is no cytotoxicity, skin irritation, or sensitization reaction;

4. The product has been clinically validated for 7 years, with sensitive measurement and precise numerical accuracy;

5. Good product compatibility

6. Strong anti-interference ability, the sensor has a certain degree of anti-interference ability against the electromagnetic signals of other electrical equipment.